Troubleshooting a Luminex® MAGPIX® Calibration & Verification Failure

Troubleshooting a Luminex® MAGPIX® Calibration & Verification Failure Running a weekly Calibration and daily performance verification on the MAGPIX® instrument is essential for maintaining instrument performance. Calibration and performance verification checks the accuracy of your MAGPIX results by ensuring the MAGPIX instrument is placing beads into the correct regions on the bead map. If the calibration and performance verification fails, it is important to troubleshoot the failure in order to get the MAGPIX instrument working correctly again. Often, MAGPIX calibration and performance verification failures occur because of low bead count, incorrect target values, reagent issues, probe problems, clogs, or debris in the imaging chamber. Let’s go through the steps to troubleshoot a calibration and performance verification failure on the MAGPIX instrument and get your instrument up and running again. First, let's check the calibration and verification report then clean the probe. MAGPIX Calibration & Verification Troubleshooting Troubleshooting Introduction & Probe Cleaning (1 of 3):    • Troubleshooting a Luminex® MAGPIX® Calibra...   Enhanced Startup Routine (2 of 3):    • Video   Follow Up Clean Routine (3 of 3):    • Video   https://www.luminexcorp.com/magpix-sy...