Fluke ESA 612 with Patient Leakage Tests - Electrical Safety Analyzer

In this vide we cover how to use a Fluke ESA 612 electrical safety analyzer. We demonstrate ground wire resistance testing, touch currents tests, and patient leakage tests. The test performed are in line with NFPA 99 - AAMI/ANSI 99 testing standards Ground wires resistance values should be less than 0.5 ohms Touch current should be run in No Fault Condition (ground intact) with the DUT in the on and off state less than 100 microamps Single Fault Condition (ground lifted) with the DUT in the on and off state less than 500 microamps Patient Leakage and Patient Auxiliary Leakage No Fault Condition (ground intact) with the DUT in the on and off state less than 100 microamps Single Fault Condition (ground lifted) with the DUT in the on and off state less than 500 microamps Hope you enjoy the video make sure to subscribe to learn more! Some of my favorite Fluke Products! Voltage Detector - https://amzn.to/3rfapv9 Fluke Multimeter and Clamp Combo Kit - https://amzn.to/3xddAax Fluke VT04A Infrared Imager - https://amzn.to/3E866Y1